Digital Radiography Articles

Reducing the Radiation Dose During Excretory Urography: Flat-Panel Silicon X-ray Detector Versus Computed Radiography

 

OBJECTIVE
The purpose of the study was to examine the possibilities for reducing radiation exposure in uroradiology using digital flat-panel silicon X-ray detector radiography. We compared the subjectively determined image quality of abdominal radiographs and urograms obtained on a digital flat-panel detector radiography system with those obtained on a computed radiography system.

 

SUBJECTS AND METHODS
Fifty patients who had a clinical indication for urography underwent unenhanced abdominal imaging that was alternately performed using flatpanel silicon X-ray detector radiography or computed radiography. For patients who required a second radiograph with contrast medium, the examination modality was changed to avoid exposing the patients to excess radiation. The images obtained on flat-panel X-ray detector radiography were obtained at half the radiation dose of the images obtained on computed radiography (800 speed vs 400 speed). The resulting 50 pairs of images were interpreted by four independent observers who rated the detectability of structures of bone and the efferent urinary tract relevant to diagnosis and compared the image quality.

 

RESULTS
At half the radiation dose, digital flat-panel X-ray detector radiography provided equivalent image quality of the liver and spleen, lumbar vertebrae 2 and 5, pelvis, and psoas margin on abdominal radiographs. The image quality obtained with digital flat-panel X-ray detector radiography of the kidneys, the hollow cavities of the upper efferent urinary tract, and the urinary bladder was judged to be statistically better than those obtained with computed radiography.

 

CONCLUSION
With half the exposure dose of computed radiography, the flat-panel X-ray detector produced urograms with an image quality equivalent to or better than computed radiography.


1 M. Zähringer, V. Hesselmann, O. Schulte, B. Krug, K. Lackner
2 K. F. Kamm
3 W. Braun
4 G. Haupt1

 

 1Department of Radiology, University of Cologne, Medical School, Joseph-Stelzmann-Str. 9, 50924 Cologne, Germany. Address correspondence to M. Zähringer
 2Philips Medical Systems, Röntgenstraße 24, 22335 Hamburg, Germany
 3Department of Medical Statistics, Informatics and Epidemiology, University of  Cologne, Medical School, 50924 Cologne, Germany
 4Department of Urology, University of Cologne, Medical School, 50924 Cologne, Germany.

AJR 2003;181:931–937
0361–803X/03/1814–931
© American Roentgen Ray Society

For full article click here:American Journal of Roentgenology.

 

©2004-2009 Koninklijke Philips Electronics N.V. All rights reserved.